CanalPro
CL2 Cordless Handpiece
TM
4.6 Insert or Remove NiTi File
4.6.1 While pressing and holding the the button on the back of the the contra
angle, lightly rotate and insert the file.
4.6.2 Ensure the file is correctly installed by lightly pulling on the file to ensure it
does not come loose.
4.6.3 A file is removed by pressing the button and pulling the file while the button
is pressed.
CAUTION!
4.7
Buzzer Setting
4.7.1 The buzzer can be set for On/Off when the preset torque reading has
12
•
Ensure the power is OFF before inserting/removing a file.
•
Check that the file is properly installed by lightly pulling
on the file.
•
Clean the file prior to inserting to avoid improper
insertion.
•
Do not exceed the recommended RPM for the file in use.
•
Failure to comply with these Cautions may result in
equipment damage or injury to personnel.
been reached.
4.7.1.1 The buzzer select menu is displayed by pressing the
power switch while pressing and holding the select switch
after turning the power Off. When the buzzer indicator is
displayed, a warning sound is made unless the buzzer is set
to Off.
4.7.1.2 The warning sound can be alternately selected (on/off) by
pressing the select switch while pressing and holding the
power switch.
4.7.1.3 The warning on/off selection can be stored by releasing the
power switch and can be used for general conditions.