Leveling; Checking The Weighing Platform - Mettler Toledo PBD659 Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 69

3.3 Leveling

Only a weighing platform which is aligned exactly horizontally supplies exact weighing results. The weighing
platform has to be leveled during the initial installation and whenever its location is changed.
1 Turn the adjustable feet of the weighing platform until the air
bubble of the spirit level is positioned in the inner circle.
2 Tighten the lock nuts of the adjustable feet.
Scan the QR code to watch the video

3.4 Checking the Weighing Platform

Function Check
1 Make sure that the weighing platform is connected to a
weighing terminal and that the weighing terminal is
switched on.
2 Make sure that the weighing platform is unloaded and
the display of the weighing terminal shows 0.
3 Load the weighing platform. The weighing terminal
must show a value different from 0.
4 Unload the weighing platform. The weighing terminal
must return to 0.
Verification Test
For a verification test refer to the user manual of the connected weighing terminal. If the verification seal is
broken, verification is no longer valid.