Identification of Symbols
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Product is intended for use in the electromagnetic environment specified below.
The customer or the user of the Product should assure that it is used in such an
environment.
Immunity test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transients /
bursts,
IEC 61000-4-4
Surges
IEC 61000-4-5
42/645
IEC 60601-1-2
Test level
± 8 kV contact
± 8 kV contact
± 15 kV air
± 15 kV air
± 2 kV for
± 2 kV for
power supply
power supply
lines
lines
± 1 kV for
Not Applicable
input/ output
lines
± 1 kV Line-to-
± 1 kV Line-to-
line
line
± 2 kV Line-to-
± 2 kV Line-to-
ground
ground
Compliance
level
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Mains power quality should be
that of a typical public low-
voltage power supply network
that supplies buildings used
for domestic purposes,
commercial or hospital, clinic
environment.
Mains power quality should be
that of a typical public low-
voltage power supply network
that supplies buildings used
for domestic purposes,
commercial or hospital, clinic
environment.
END-IFU-PROPEX-PIXI-V15-WEB-EUR-Multilingual-2021-10
Electromagnetic
environment - guidance