Operation; Control Unit - Nouvag TCM 3000 Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

TCM 3000

5.1 Control unit

Starting the unit
Turn green power switch "ON/OFF" to (ON) position(on rear of unit)
Display:
The display will show the current speed and torque of the instrument.
Keyboard panel:
„Motor":
Switch motor ON/OFF (the light will illuminate when the motor is switched on)
„▲":
Press to increase speed or torque
„▼":
Press to decrease speed or torque
To change the direction of rotation of the motor → on reverse mode the control unit
„Rev/Forw":
gives an acoustical signal.
„1:1":
For use with 1:1 contra angle
Speed: 2,000 to 30,000 rpm
Torque AL can be selected within the range: 1, 2, 3, 4, 5 Ncm
AS- Stop modus does not function
„16:1":
For use with 16:1contra angle
Speed: 125 to 1,875 rpm
Torque AL can be selected within the range: 10, 15, 20, 25, 30 Ncm
AS-Stop modus does not function
„70:1":
For use witth 70:1contra angle
Speed: 15 to 440 rpm
Torque AL can be selected within the range: 10, 15, 20, 25, 30, 35, 40, 45, 50, 55 Ncm
Toruqe AS can be selected within range: 10, 15, 20, 25, 30, 35, 40, 45, 50, 55 Ncm (up
to 60 rpm only)
F F F F
„ATC":
To select (Automatic Torque Control) AL- and AS function
The key „Motor" functions only when the pedal is not connected.
4
All manuals and user guides at all-guides.com

5 Operation

NOUVAG AG
No. 31891 31/05

Publicidad

Tabla de contenido
loading

Tabla de contenido