6. Commissioning, operation
3) The output signal can be adjusted to the set point by means of the knob. For this
purpose select the decade to be set using the arrow keys.
4) The output signal can also be increased incrementally in steps of 4 mA (25 %) using
GB
the "25%" key.
5) The automatic step function and automatic ramp function allow automatic change of
the output signal.
For further information refer to the section "6.3.1 Automatic operation modes" of the
instruction.

6.2.2 Milliampere simulation

Figure 4 shows the way the CEP1000 is connected if it is supposed to function as a
2-wire transmitter with an external power supply. In this case the circuit is powered via
an external voltage source and the instrument just regulates the current in the circuit.
The regulating procedure of the output signal is the same as for the application as the
source of current in the milliampere range.
SIM
LOOPPWR
Calibration Line
Figure 4: Simulation of a sensor with a current output for 4 ... 20 mA
16
4 ... 20 mA
CEP 1000
mA
MAX
MODE
AUTO
MIN
%ERROR
25%
The instrument is operated in the simulation mode as a
two-wire sensor.
( + )
( - )
( + )
Instrument with an input
( - )
WIKA operating instructions, model CEP1000
Power supply DC 30 V
max.
of 4 ... 20 mA