Data Acquisition For Minimum And Maximum - WIKA Obsolete CEP1000 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

6. Commissioning, operation

6.3.3 Data acquisition for minimum and maximum

If the CEP1000 is operated in a measuring mode (milliampere, milliampere with power
supply or voltage), the minimum and maximum values for the corresponding input mode
GB
are monitored and stored permanently. To activate the data acquisition proceed as
follows:
1) Switch the CEP1000 to a measuring mode.
2) Connect the instrument to the instrument to be tested.
3) Press both keys ("MIN" and "MAX") to clear the memory for the minimum and
maximum values. The "CLR" message is displayed on the LCD display, if the
memory has been deleted.
4) Acquire the minimum value and the maximum value using the instrument as long as
required.
5) The saved value can be accessed any time by pressing the "MIN" or "MAX" key.
6) When switching off the instrument or switching to another operating mode the
memory will be cleared. Back up the data before switching off.
20
If necessary, deactivate the battery saving function prior to calling the
data acquisition function to avoid premature switching off of the calibra-
tor.
WIKA operating instructions, model CEP1000

Publicidad

Tabla de contenido
loading

Tabla de contenido