Optional Automatic Wash / Disinfect
CAUTIONS
• Alkaline and highly acidic detergents and solutions
cause damage to the devices.
• Channels and crevices found on these devices require
particular attention during cleaning.
• Pay special attention to the water quality used
throughout reprocessing. Hard water can damage
the surface of the equipment. Avoid using hard water.
Instead use purified water unless otherwise specified.
Limitations on Reprocessing
• Repeated processing has negative effects on these
devices and is not recommended for routine use.
• It is important to have Integra NeuroSpecialists
perform routine inspections (twice yearly is
recommended).
See contact information.
INSTRUCTIONS
Cleaning - Automated Equipment: Neutral pH Detergent
Method
1. Prepare equipment for cleaning by disassembling
removable parts and loosening connections.
2. Rinse equipment in warm water before placing into
washer.
3. Load device into the washer and place small parts in
container or tray inside the washer unit in order to
avoid losing small components.
NOTE – Load devices carefully into washer in order to
avoid collision.
4. Follow the instructions listed below and set washer
machine to these exact parameters:
Phase
Time
Water Tem-
(Min.)
perature
Pre-wash
4:00
Cold water
1
Enzyme
4:00
Hot water
Wash
Wash 1
10:00
60.0ºC
(140º F)
Rinse 1
0:30
Hot water
Thermal
2:00
82.2ºC
Rinse**
(180º F)
Detergent and
Concentration
N/A
Neutral pH
enzymatic (e.g.
Endozime® AW
Triple Plus with
APA, Ratio 1:128)
Neutral pH
detergent
(e.g. Renu-Klenz™
, Steris, Ratio
1:256)
N/A
N/A
7
** Optional phase for disinfection of components –
minimum water temperature as indicated or per worker
manufacturer specifications for the thermal rinse cycle.
NOTE – Any deviation from this guideline could result in
damage to the equipment as well as improper cleaning
results.
Rinse with purified water. Do not perform if parameters
cannot be achieved.
5. Remove from washer and dry completely if needed.
6. Inspect equipment to make sure there is no visible
organic debris or residue from cleaning agent.
Repeat process if any visible soil is detected.
The life expectancy of the MAYFIELD products is expected
to be as long as 7 years.
Maintenance and Care
To ensure proper function and to extend the life and
performance of the equipment, Integra LifeSciences
recommends the following:
Recommended Action
Return the device to
the Integra LifeSciences
Repairs department for
detailed inspection and
servicing.
Request that Integra
NeuroSpecialists perform
routine inspections of the
device
In the absence of proper care and servicing of the device,
negative effects may be seen after repeated processing
over time which may lead to reduced performance.
Contact information: See the Service and Repair section
for contact information on how to return your device for
periodic servicing and to request periodic inspections.
See Inspection and/or Service notes section for routine
checks to be performed on the device.
NOTE: Any serious incident that has occurred in relation
to the device for the user and/or the patient should
be reported to the manufacturer and the competent
authority of the member state in which the user and/or
patient is established.
EN – ENGLISH
Recommended Frequency
Once / year
Twice / year