12. TECHNICAL SPECIFICATIONS

Type
Model no.
Dimensions (L x W x H)
Weight
Light
Power output
CRI
Light intensity
Radiation
Operating conditions
Transport and storage conditions
Product classification
Accessories
Colour temperature of the LEDs
Threshold of short wavelength
Maximum light output
Expected service life of the device 10.000 hours
Maximum radiance output of the TL 85
Radiance output
E
: Eye UV-A
UVA
ES: Actinic UV skin & eye
EIR: Infrared radiation hazard ex-
posure limits for the eye
LIR: Retinal thermal
(mild visual irritation)
LB: Blue light
LR: Retinal thermal
Subject to technical changes.
The serial number is located on the device or in the battery compartment.
Brightness: 10,000 lux (this point about the light intensity is merely for information purposes. With regard
to the standard IEC 60601-2-83, this light source is classified as an Exempt Group).
GCE504
TL 85
30 x 23.5 x 45.2 cm
approx. 1340 g
LEDs
max. 60 watts
min. 90
12,000 lux (at a distance of approx. 20 cm)
Radiance output outside of the visible range (infrared and UV) is
low enough that no eye or skin damage is anticipated.
0°C to +35°C, 15 - 90% relative humidity,
700 – 1060 hPa ambient pressure
-20°C to +60°C, 15 - 90% relative humidity,
700 – 1060 hPa ambient pressure
Protection class II, IP21
Mains adapter, instructions for use, stand
6500K +/-400 Kelvin
460 - 670 nm
52.27 W / m²
Risk group classified in ac-
cordance with IEC 62471
Exempt Group
Exempt Group
Exempt Group
Exempt Group
Exempt Group
Exempt Group
22
Maximum value
4.393e-5
2.794e-6
4.631e-1
1.491e-1
7.057e0
9.327e1