SPECIFICATIONS

Catalog Number
Delivery Rate (Lower delivery rates available
for low flow applications)***
Maximum Recommended Flow (@ nominal
outlet pressures of zero & 7 kPa)**
Outlet Pressure
Auxiliary Oxygen Port **
Electrical Rating
Operating Voltage Range
Oxygen Percentage
Operating Atmospheric Pressure
1010 hPa to 840 hPa
0-1500 M (0-4921 ft)
840 hPa to 616 hPa
1500-4000 M (4921-13123 ft)
Operating Temperature Range
Operating Relative Humidity Range
Power Consumption
Weight
Safe Working Load
Sound Pressure Level at 3 and 5 LPM
(ISO 80601-2-69)
Sound Power Level at 3 and 5 LPM
(ISO 80601-2-69)
Sound Level (ISO 8359:1996)
Alarm Sound Level
Dimensions
Maximum Limited Pressure
Operating System
Low Oxygen Indicator
Storage Conditions
Equipment Class and Type
Approval Body and Safety Standard
CE mark
EMC Compliance To
**
CAUTION – The maximum recommended flow is 3 LPM when an oxygen bottle is being filled with oxygen from the auxiliary oxygen port.
***
NOTE – Use of the 515LF-607 low output flow meter package or other low output flow meter accessory will prevent the device from meeting the requirements of ISO-80601-2-69:2014
Section 201 .13 .2 .101 .
Specifications subject to change without notice.
A-525DS
DEVILBISS 5 - LITER SERIES
525DS, 525DS-Q
0 .5 to 5 LPM
5 LPM
8.5 ± 0.5 psig (58.6 ± 3.5 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
115 V, 60 Hz, 3 .3 Amp
97-127 V~, 60 Hz
1-5 LPM=87%-96%
Across the voltage range:
No degradation of performance
Tested at nominal voltage only:
No degradation of performance
41˚F (5˚C) to 95˚F (35˚C)
15% to 93%, non-condensing
310 Watts Average
275 Watts @ 1.2 LPM & below
36 lbs. (16.3 Kilograms)
53 lbs. (24 Kilograms)
525DS
525DS-Q
50.9 dBA @ 3 LPM
46.7 dBA @ 3 LPM
50.7 dBA @ 5 LPM
46.7 dBA @ 5 LPM
525DS
525DS-Q
54.7 dBA @ 3 LPM
50.4 dBA @ 3 LPM
54.5 dBA @ 5 LPM
50.4 dBA @ 5 LPM
48 dBA (525DS)
46 dBA (525DS-Q)
> = 62 dBA
24 .5"H x 13 .5"W x 12"D
(62.2 x 34.2 x 30.4 cm)
Normal Condition: 9 PSIG (62.0 kPa),
Single Fault Condition: 27.6 PSIG (190.3 kPa)
Time Cycle / Pressure Swing
<82% low oxygen
<60% very low oxygen
-13˚F (-25˚C) to 158˚F (70˚C), humidity range
of 15% to 93% non-condensing
Class II Equipment Double Insulated;
Type B Applied Part, IP21
TUV
ANSI/AAMI ES60601-1:2005+A2 (R2012) +A1
IEC 60601-1-6:2010
IEC 60601-1-11:2015
ISO 80601-2-69:2014 ***
CAN/CSA-C22 .2 No . 60601-1:14
CAN/CSA-C22 .2 No . 60601-1-6:11
CAN/CSA-C22 .2 No . 60601-1-11:15
CAN/CSA-C22 .2 No . 80601-2-69:16
No
EN60601-1-2
525KS, 525KS-LT
0 .5 to 5 LPM
5 LPM
8.5 ± 0.5 psig (58.6 ± 3.5 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
220-230 V~, 50 Hz, 1 .55 Amp
230 V~, 60 Hz, 1 .9 Amp
187-253 V~, 50 Hz
195-253 V~, 60 Hz
1-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
Tested at 230V/50Hz only:
No degradation of performance
41˚F (5˚C) to 95˚F (35˚C)
15% to 93%, non-condensing
230V / 50 Hz - 312 Watts Average
230V / 50 Hz - 296 Watts Average @
1 .2 LPM & below
230V / 60 Hz - 387 Watts Average
230V / 60 Hz - 369 Watts Average @
1 .2 LPM & below
36 lbs. (16.3 Kilograms)
53 lbs. (24 Kilograms)
525KS
525KS-LT
47.9 dBA @ 3 LPM
49.6 dBA @ 3 LPM
47.9 dBA @ 5 LPM
49.4 dBA @ 5 LPM
525KS
525KS-LT
51.6 dBA @ 3 LPM
53.4 dBA @ 3 LPM
51.7 dBA @ 5 LPM
53.2 dBA @ 5 LPM
40 dBA (50 Hz (525KS)
48 dBA (50 Hz (525KS-LT)
> = 62 dBA
24 .5"H x 13 .5"W x 12"D
(62.2 x 34.2 x 30.4 cm)
Normal Condition: 9 PSIG (62.0 kPa),
Single Fault Condition: 27.6 PSIG (190.3 kPa)
Time Cycle / Pressure Swing
<82% low oxygen
<60% very low oxygen
-13˚F (-25˚C) to 158˚F (70˚C), humidity range
of 15% to 93% non-condensing
Class II Equipment Double Insulated;
Type B Applied Part, IP21
TUV approved for 50 Hz only to
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
Yes
EN60601-1-2
525PS
0 .5 to 5 LPM
5 LPM
8.5 ± 0.5 psig (58.6 ± 3.5 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
220-230 V~, 60 Hz, 1 .68 Amp
187-253 V~, 60 Hz
1-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
Tested at 230V/60Hz only:
No degradation of performance
41˚F (5˚C) to 95˚F (35˚C)
15% to 93%, non-condensing
230V / 60 Hz - 334 Watts Average
230V / 60 Hz - 297 Watts Average @
1 .2 LPM & below
36 lbs. (16.3 Kilograms)
53 lbs. (24 Kilograms)
45.4 dBA @ 3 LPM
45.3 dBA @ 5 LPM
49.2 dBA @ 3 LPM
49.1 dBA @ 5 LPM
> = 62 dBA
24 .5"H x 13 .5"W x 12"D
(62.2 x 34.2 x 30.4 cm)
Normal Condition: 9 PSIG (62.0 kPa),
Single Fault Condition: 27.6 PSIG (190.3 kPa)
Time Cycle / Pressure Swing
<82% low oxygen
<60% very low oxygen
-13˚F (-25˚C) to 158˚F (70˚C), humidity range
of 15% to 93% non-condensing
Class II Equipment Double Insulated;
Type B Applied Part, IP21
TUV
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
Yes
EN60601-1-2
EN - 11