Devilbiss Osd ® Operation; Reserve Oxygen System; Exterior Cabinet - DeVilbiss 525 Serie Guía De Instrucciones

Concentrador de oxígeno de 5-litros
Ocultar thumbs Ver también para 525 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

DANGER
Oxygen makes it easier for a fire to start and spread. Do not leave the nasal cannula or mask on bed coverings or chair cushions if the oxygen concentrator is
turned ON but not in use, the oxygen will make the materials flammable. Turn the oxygen concentrator OFF when not in use to prevent oxygen enrichment.
NOTE – If the audible signal alarms but the unit is not operating, there is no power to the unit. Refer to the Troubleshooting chart on page 10, and contact your
DeVilbiss provider if necessary.
NOTE – If an audible low-frequency vibration sound is detected, the unit is not operating properly. Refer to the Troubleshooting chart on page 10, and contact your
DeVilbiss provider if necessary.
2.
Check the flow meter to make sure that the flow meter ball is centered on the line next to the prescribed number of your flow rate.
CAUTION
It is very important to follow your oxygen prescription. Do not increase or decrease the flow of oxygen – consult your physician.
NOTE – Your DeVilbiss provider may have preset the flow meter so that it cannot be adjusted.
NOTE – If the flow meter knob is turned clockwise, the flow decreases (and eventually will shut off the oxygen flow). If the knob is
turned counterclockwise, the flow increases.
NOTE – For prescriptions of 5 LPM, be sure the ball is centered on the 5 liter line. The ball should not touch the red line. Setting the
flow higher than 5 may cause the oxygen purity level to drop.
NOTE – The low-flow alarm may activate if the flow meter ball is set at or below 0.2 lpm. The unit will continue to run; however, the
Service Required light will come on accompanied by an audible alarm. Adjust the flow meter to your prescribed flow.
3 .
Your DeVilbiss concentrator is now ready for use, properly position the cannula with the nasal prongs facing upward, insert the prongs
into nose. Wrap the cannula tubing over the ears and position in front of body (Figure 4). Allow 15 minutes for the oxygen concentrator
to reach stated performance .
DeVilbiss OSD
Operation
®
The OSD (Oxygen Sensing Device) is a device within your concentrator that monitors the oxygen produced by your unit.
The OSD lights on the top panel are defined as follows:
Green Normal Oxygen light–acceptable oxygen level.
Yellow Low Oxygen light– below an acceptable oxygen level.
If the oxygen purity falls below the acceptable level: The green Normal Oxygen light will shut off, the yellow Low Oxygen light will illuminate, and an intermittent audible signal
will sound .
Refer to the Troubleshooting section in this guide on page 10, and switch to your reserve oxygen system . Do not attempt any other maintenance . Contact your DeVilbiss
provider immediately .

RESERVE OXYGEN SYSTEM

As a precaution, your DeVilbiss provider may supply you with a reserve oxygen system . If your unit loses electrical power or fails to operate correctly, the Patient Alert System
will sound to signal you to switch to your reserve oxygen system (if provided) and contact your DeVilbiss provider. Please contact your oxygen provider if you have questions
regarding a reserve oxygen system .
CARING FOR YOUR DEVILBISS OXYGEN CONCENTRATOR
DeVilbiss recommends using only original DeVilbiss parts and filters in order to guarantee reliable operation of the product.
WARNING
Do NOT use lubricants, oils or grease .
Before attempting any cleaning procedures, turn the unit "OFF ."
Cannula/Mask, Tubing and Humidifier Bottle
Clean and replace the cannula/mask, tubing and humidifier bottle according to the manufacturer's instructions.
Filter Door with Vents
Inspect the vents periodically, and wipe with a dry cloth as needed to remove dust .

Exterior Cabinet

Clean the concentrator exterior cabinet weekly by using a damp cloth and wiping dry; the vents can also be wiped with a damp cloth .
Cleaning
Outer Cabinet
Filter Door Vents
Oxygen Outlet Connector
* number of cleaning cycles determined by recommended cleaning interval and expected service life
WARNING
To avoid electric shock, do not plug the concentrator into an AC outlet if the concentrator cabinet is broken . Do not remove the concentrator cabinet . The cabinet
should only be removed by a qualified DeVilbiss technician. Do not apply liquid directly to the cabinet or utilize any petroleum-based solvents or cleaning agents.
CAUTION
Use of harsh chemicals (including alcohol) is not recommended. If bactericidal cleaning is required, a non-alcohol based product should be used to avoid
inadvertent damage .
A-525DS
Recommended cleaning interval
7 days
7 days
7 days
Number of cleaning cycles *
260
260
104
4
Compatible cleaning method
Water, use only a damp cloth
Wipe with dry cloth, or a cloth dampened with
water to remove dust .
Mild dish soap (2 tbsp) and warm water (2 cups)
EN - 9