6 Limitations on Use

Only for use with BladeXT in conjunction with a Blatchford approved pyramid
system.
Intended Life
A local risk assessment should be carried out based upon activity and usage.
Lifting Loads
User weight and activity is governed by the stated limits.
Load carrying by the user should be based on a local risk assessment.
Environment
The product is waterproof to a maximum depth of 1 metre.
Thoroughly rinse with fresh water after use in abrasive environments such as those that may
contain sand or grit, for example, to prevent wear or damage to moving parts.
Thoroughly rinse with fresh water after use in salt or chlorinated water.
Exclusively for use between -15 °C and 50 °C
(5 °F to 122 °F).
.

7 Fitting Advice

The instructions in this section are for practitioner use only.
• A recurring noise occurs between BladeXT
and the adapter.
• The adapter moves out of position.
Symptoms
1. All bolts and nuts must be tightened to the
torque ratings specified in this document.
2. Apply Loctite where applicable.
3. User must not use the device until
adjusted, repaired or replaced.
7
Remedy
938344PK1/1-0421