2. Remove the three Phillips-head screws from the copper anti-fouling assembly, and pull the
copper guard off of the plastic anti-fouling holder.
3. Remove the protective plug from the anti-fouling device cup.
4. Use a toothpick to lift each of the anti-fouling devices out of the holder. If necessary, use needle-
nose pliers to carefully break up the device
Option
To deploy sensor
To clean or store
sensor
Section 4 General information
This product can expose the user to chemicals with silica, crystalline (airborne particles of respirable size), which
is known to the State of California to cause cancer and birth defects or other reproductive harm. For more
information, go to www.P65Warnings.ca.gov.
Revised editions of this user manual are on the manufacturer's website.

4.1 Service and support

The manufacturer recommends that sensors be sent back to the manufacturer annually to be
cleaned, calibrated, and for standard maintenance.
Refer to the website for FAQs and technical notes, or contact the manufacturer for support at
support@seabird.com. Do the steps below to send a sensor back to the manufacturer.
1. Complete the online Return Merchandise Authorization (RMA) form or contact the manufacturer.
Note: The manufacturer is not responsible for damage to the sensor during return shipment.
2. Remove all batteries from the sensor, if so equipped.
Procedure
Insert new anti-fouling devices into the cup, then install the cap onto the cup. Do not
tighten too tight.
Attach the copper assembly to the sensor again.
Do not insert new anti-fouling devices. Install the protective plug. Make sure to remove
the plug before the next deployment or pressurization of the sensor. Damage to the
conductivity cells can be caused if the plugs are not removed.