CALIBRATION
This scale has been factory calibrated, and does not require calibration prior to use.
If required, the scale can be calibrated. Calibration of your scale is performed using kilograms
(KG) or pounds (LB), according to the units of measure used upon entering into the calibration
path. When calibrating in pounds, a total of 300 lbs is required., applied in increments of 100 lb,
200 lb, and 300 lb. If calibrating in kilograms, a total of 120 kg are required , applied in
increments of 40 kg, 80 kg, and 120 kg.
Note: Only weights certified and traceable to national standards are to be used for
calibration procedures.
With the scale off, press and hold the LB/KG button (499KL) or the
blank blue button (499KG) and ZERO buttons at the same time, then
press the ON/OFF button to turn on the scale. Release all three
buttons when "CAL" appears on the display. Numbers will appear
indicating that you can begin calibration.
When calibration numbers stabilize, press the ZERO button. "C-100"
(for LB mode) or "C-40" (for KG mode) will flash on the display,
followed by "0" or "1".
Place 100 lbs or 40 kg on the scale. Numbers will appear on the
display. When calibration numbers stabilize, press the ZERO button.
"C-200" (for LB mode) or "C-80" (for KG mode) will appear on the
display, followed by "0" or "1".
Place an additional 100 lbs or 40 kg on the scale, for a total of 200
lbs or 80 kg. Numbers will appear on the display. When calibration
numbers stabilize, press the ZERO button. "C-300" (for LB mode) or
"C-120" (for KG mode) will appear on the display, followed by "0" or
"1".
Place an additional 100 lbs or 40 kg on the scale, for a total of 300
lbs or 120 kg. Numbers will appear on the display. When calibration
numbers stabilize, press the ZERO button. The calibration procedure
is now complete and "END" will appear on the display followed by
"OFF". The scale will automatically shut off.
UM499KL REV 20210305
ACTION
80
Back to Home Page
DISPLAY SHOWS
"CAL" followed by
numbers
Numbers followed
by "C-100" or
"C-40" then "0" or
"1"
Numbers followed
by "C-200" or
"C-80" then "0" or
"1"
Numbers followed
by "C-300" or
"C-120" then "0"
or "1"
"END" then "OFF"