Greengate
Greengate Patient Control Station and
Receptacle (GPCS/GG37P)
Installing the GPCS and GG37P
Mount the GG37P on the headwall to a single gang wall box with a minimum internal depth of 2 inches (51mm). The GG37P
receptacle contains three bundles of cabling. Cable A is typically used for connection to the Room Controller. For GPCS
models with dimming raise/lower controls, an additional three wires will be connected from Cable C connections. All other
connections on the GG37P are for termination to the Nurse Call and Entertainment system.
Pillow Speaker
Connection Point
Switchpack
White
Receptacle
Red
BMS/Out
Black
Sensors
Green
Sensors
Slider
Station
Wallstations
-
A/V
+
Mode
-
Demand
+
Response
-
Alert
+
Mode
-
Time
+
Clock
Energy Options DIP Switch
Demand Response
Occupancy
Not Used
1 2
3
4
Default 10%
Occ
20%
Vac (default)
30%
40%
0-10V Dimming Outputs
-
-
-
+
+
+
Dimmer 3
Dimmer 2
Dimmer 1
Occupancy/Vacancy
Mode Selection
Click & Go RJ45 connector for Wallstations, Slider, Sensors,
Receptacle control, Alternate Voltage Switchpack control, and BMS Output
See next page for wiring table.
N
ote:
9850-000366-04 page 7
October 2021
RED (PIN 20) All Lighting On/Off
BLK (PIN 2) Lighting Zone 3
GRN (PIN 21) Lighting Zone 2
WHT (PIN 3) Lighting Zone 1
4
3
2
1
Dimmer 3
-
Dimmer 2
-
Dimmer 1
-
+
+
+
0-10V Dimming
Three removable
connectors for 0-10V
dimming outputs
Quick connect cables are pre-terminated
and included in the Room Controller
QuicKit and measured to fit typical room layouts
Specifications
Voltage
24 VDC supplied from Room Controller
Electrical
Class 2, LPS
GPCS requires the GG37P for connection,
Connections
which connects to room controller via
provided terminal blocks
GRN/BLK (PIN 37) +24 VDC
BRN (PIN 27) +24 VDC
RED/BLK (PIN 6) Selected Lighting Lower
ORN/BLK (PIN 24) Selected Lighting Raise
Cable B
(Pins 5, 24, 27)
Cable C
Typically connects to
Nurse Call Stations
Cable A
Cable D
(Pins 2,3,20,21,27)
Typically connects to
entertainment devices
GG37P
Greengate 37 Pin Receptacle
Provided by Cooper Lighting Solutions
for pillow speaker connection
General
Exam
Reading
Raise
Lower
All Off
HC1
Health Care Station
Room Controller for Healthcare
GPCS
Greengate Patient Control Station
Provided by Cooper Lighting Solutions
Note:
Use only wiring supplied with the GG37P receptacle.
Only the Greengate Patient Control Station shall be
connected to a GG37P Receptacle
www.cooperlighting.com