Calibration Menu - OHAUS Scout SJX Serie Manual De Instrucciones

Ocultar thumbs Ver también para Scout SJX Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

EN-10
AZT locked to 0.5d
Zero range forced to 2%
Filter Level is locked to current setting.
Stable only in Print menu locked to ON
Continuous/Interval print cannot be selected (SJX...N/E models only)
Security switch:

4.2 Calibration Menu

Enter this menu to perform calibrations.
Internal Calibration*
Calibration Adjustment* -100...0...100
Span:
Linearity:
End Calibration:
Notes: * These sub menus are not available in SJX/E models. The settings vary with model.
Internal Calibration [I I n.CAL]
Initiates an internal calibration procedure.
Calibration Adjustment [CAL.ADJ]
Set the calibration adjustment factor for the internal calibration mass.
-100...0...100
Note: Before making a calibration adjustment, perform an Internal Calibration. To verify whether
an adjustment is needed, place a test mass equal to the span calibration value on the pan and
note the difference (in divisions) between the nominal mass value and the actual balance
reading. If the difference is within ± 1 division, calibration adjustment is not required. If the
difference exceeds ± 1 division, calibration adjustment is recommended.
Example:
Span [SpaN]
Initiates a span calibration procedure (zero and span)
Lin [LIN]
Initiates a linearity calibration procedure (zero, mid-point and span).
End Cal [End]
Advance to the next menu or return to the top of the current menu.
Find Quality Products Online at:
Unlocked
Figure 4-1. Security switch.
Perform
Perform
Perform
Exit menu
Actual weight reading:
Expected weight reading: 1500.00 (Test mass value)
Difference Weight (d):
Difference weight in digits: -14 (Adjust value)
GlobalTestSupply
www.
TM
SCOUT
Locked
1500.14
0.14
.com
SJX SERIES BALANCES
sales@GlobalTestSupply.com

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido