Table 3
Radiated RF
IEC61000-4-3
(Test specifications for
ENCLOSURE PORT
IMMUNITY to
RF wireless communi-
cations equipment)
Guidance and manufacturer's declaration - electromagnetic Immunity
Test
Band
Frequency
(MHz)
(MHz)
385
380 -
390
450
430 -
470
710
704 -
787
745
780
Service
Modulation
TETRA 400
Pulse
modulation
18 Hz
GMRS 460,
FM
FRS 460
± 5 kHz devi-
ation
1 kHz sine
LTE Band
Pulse
13,
modulation
17
217 Hz
110
Modulation
Distance
(W)
(m)
1.8
0.3
2
0.3
0.2
0.3
IMMUNITY
TEST LEVEL
(V/m)
27
28
9