Warnings
• Ensure that the user's feet do not get caught between the footplate and the floor. Ensure that the user's feet do not
get caught between the footplate and the base when adjusting the base width, or that a caregiver's hands do not
get caught or pinched if adjusting the user's feet on the footplate.
• Do not use the calf strap accessory without also using a sling that is appropriate for the MiniLift, such as the
ThoraxSling.
• To transfer with MiniLift200, the user must be able to support weight while standing.
To transfer with MiniLift200, the user must be able to:
• support weight while standing
• understand instructions
Safe working load
Different products on the same lift system (lift unit, sling bar, sling, scales and other lifting accessories) may have different
allowable safe working loads. The lowest allowable safe working load always determines the safe working load of the
assembled system. Always check the safe working loads for the lift and accessories before use. Contact your dealer if
you have any questions.
The control box and battery have the following features
1. Emergency stop
2. Charging lamps
3. Display showing battery charge level
4. Connection for charger cable
5. Connection for base-width adjustment motor
6. Connection for lift arm actuator
7. Connection for hand control
8. Electrical emergency lowering
9. Handle for lifting battery