d) Automatic calibration will end when the SV field stops flashing and the measured
temperature PV is equal to the set temperature SV.
Mixing speed regulation
Although the maximum mixing speed that can be set in the device is
Caution:
greater than 1800 rpm, it is not recommended to exceed this value!
The method of regulating the mixing speed differs depending on whether the heating
function is active or inactive.
When the device is operated with heating function active:
a)
Press mixing activation button [I]. Field R/MIN [D] will show the rotating
speed of the agitator.
b)
Press "SET" [E]. Editing of the last digit in the SV field will be activated (the
digit starts flashing).
c)
Press the "SET" button [E] again. The R/MIN [D] field will display the set
mixing speed and it will be possible to edit the last digit of this field (the digit
will be flashing).
d)
Press "+" or "-" [H or G], to increase or reduce mixing speed. In order to
alternate between the edited digits, press "R/AT" [F].
e)
Press "SET" [E], to complete mixing speed regulation. R/MIN field [D] will
display the current mixing speed.
When the device is operated without heating function active:
a)
Switch the device on with I/O [IV].
b)
Press mixing activation button [I]. R/MIN field [D] will display the current
mixing speed.
c)
Press"SET" [E]. The R/MIN [D] field will display the set mixing speed and it will
be possible to edit the last digit of this field (the digit will be flashing).
d)
Press "+" or "-" [H or G], to increase or reduce mixing speed. In order to
alternate between the edited digits, press "R/AT" [F].
e)
Press "SET" [E], to complete mixing speed regulation. R/MIN field [D] will
display the current mixing speed.
Overload protection
The device is equipped with a fuse (X) protecting against overload. In the
event of the safeguard is activated (the device switches off), find and remove
the cause. Subsequently, replace the fuse with a new one.