Use the Blood Pressure Monitor outside of its specified operation
temperature and humidity rating, or if stored outside of its specified
storage temperature and humidity rating. Avoid storage in direct sunlight.
Use this device in a moving vehicle; inaccurate measurements may result.
Use third party accessories. Only use Microlife authorized accessories,
such as cuffs or AC adapters, as those not approved for use with the
device may provide inaccurate measurements, injury, and/or damage
the device.
STANDARDS
This medical device is compliant with:
• Medical device and non-invasive blood pressure monitor standards
IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and AAMI/ANSI/IEC
80601-2-30, and
• Electromagnetic standards IEC 60601—1-2, along with FCC Part 15, and
• Clinical Testing per standard ISO 81060-2:2013 was conducted on
blood pressure device using the same measurement technology.
Please note: According to international standards, your monitor should be
checked for accuracy every 2 years.
TYPE
IP20:
Protected against solid foreign particles with a diameter of
more than 12.5 mm, no protection against water.
Keep Dry.
Type BF Applied Part
Expected Life
Monitor
5 Years
Cuff
2 Years
Batteries and electronic instruments must be disposed of in accordance
with the locally applicable regulations, not with domestic waste.
4