CLEANING AND DISINFECTION
The product components must be cleaned thoroughly
immediately after each use and disinfected at least once a day.
4.1 Preparation
• Detach the tube from the nebuliser.
• Detach the mouthpiece from the nebuliser.
• Make sure that all residual volume is removed from the
nebuliser.
• Dismantle the nebuliser into its individual parts.
• Carefully pull the blue exhala-
tion valve out of the slot in the
mouthpiece. The valve must
still be attached to the mouth-
piece.
4.2 Cleaning
The connection tube cannot be cleaned or disinfected.
Regarding care of the connection tube, follow the
instructions in the corresponding section [see: Care of the
connection tube, page 23].
• Briefly rinse all parts used in running drinking water
beforehand.
• Place all disassembled com-
ponents in warm drinking water
with a little dishwashing liquid
for about 5 min.
• Rinse off all parts thoroughly in drinking water.
• Shake the water out of all of the parts.
– 21 –