TESTING IN CAT III / CAT IV MEASUREMENT LOCATIONS
Ensure the test lead shield is pressed firmly in place. Failure to use
the CAT III / CAT IV shield increases arc-flash risk.
5/32"
(4 mm)
TESTING IN CAT II MEASUREMENT LOCATIONS
CAT III / CAT IV shields may be removed for CAT II locations. This
will allow testing on recessed conductors such as standard wall
outlets. Take care not to lose the shields.
AC VOLTAGE (LESS THAN 600V)
1. Insert RED test lead into VΩ jack
COM jack
highest V AC (
2. Measure voltage and rotate the Function Selector Switch
to successively lower V AC (
resolution measurements.
NOTE: The hazardous voltage indicator
NOTE: Do not attempt to measure more than 600V or 200mA.
Black lead
OPERATING INSTRUCTIONS
, and rotate Function Selector Switch
4
) setting (600V).
Red lead
0.7" (18 mm)
, and BLACK test lead into
5
) settings to obtain higher
will appear for voltages >30V.
to the
2
7