MD 9231 Industrial TRMS AC/DC Current Clamp Meter
Capacitance
RANGE
200,0 μF, 2500 μF
Accuracies with film capacitor or better.
1)
Diode Tester
RANGE
2,000 V
Test Current: 0,3 mA typically.
Open Circuit Voltage: < 3,5 VDC typically.
AmpTip
clamp-on DCA
TM
RANGE
00,00 A ~ 20,00 A
20,00 A ~ 60,00 A
Induced error from adjacent current-carrying conductor: < 0,02 A/A.
1)
Specified with DC-Zero mode applied to offset the non-zero residual readings, if any.
2)
Add 10d to the specified accuracy @ < 4 A.
3)
Regular Clamp-on DCA
RANGE
60,00 A
, 600,0 A, 1000 A
3)
Induced error from adjacent current-carrying conductor: < 0,02 A/A.
1)
Specified with DC-Zero mode applied to offset the non-zero residual readings, if any.
2)
Add 10d to the specified accuracy @ < 9 A.
3)
AmpTip
clamp-on DC+ACA
TM
RANGE
DC, 40 Hz ~ 100 Hz
00,00 A ~ 20,00 A
20,00 A ~ 60,00 A
100 Hz ~ 400 Hz
00,00 A ~ 20,00 A
20,00 A ~ 60,00 A
Induced error from adjacent current-carrying conductor: <0,08 A/A.
1)
Specified with DC-Zero mode applied to offset the non-zero residual readings, if any.
2)
Add 10d to the specified accuracy @ < 4 A
3)
Regular Clamp-on DC+ACA
RANGE
DC, 40 Hz ~ 100 Hz
60,00 A
, 600,0 A, 1000 A
3.
100 Hz ~ 400 Hz
60,00 A
, 600,0 A, 1000 A
3.
Induced error from adjacent current-carrying conductor: < 0,08 A/A.
1)
Specified with DC-Zero mode applied to offset the non-zero residual readings, if any.
2)
Add 10d to the specified accuracy @ < 9 A.
3)
Maximum Crest Factor < 1.4 : 1 at full scale & < 2.8 : 1 at half scale.
4)
Accuracy
1)
2,0% + 4d
Accuracy
1.5% + 5d
Accuracy
1) 2) 3)
1,5% + 5d
3,0% + 5d
Accuracy
1) 2)
1,8% + 5d
Accuracy
1) 2) 3)
2.0% + 7d
3,0% + 7d
2,2% + 7d
3,0% + 7d
Accuracy
1) 2)
2,2% + 7d
4.
2,5% + 7d
4.
19