1.4.3 Improper maintenance, modifi cations and repairs
The subsequent installation of additional devices from other manufacturers is not permitted, nor is the modifi cation of protective devices
without consulting CADstar Technology or the manufacturer of the respective additional device.
Repairs to the scanner may only be performed by CADstar Technology customer service.
1.5 Intended use
The intended use of the scanner is the three-dimensional, optical measurement of duplicated human structures in combination with the
supplied software.
Only impressions and plaster models of the jaws, as well as articulators with plaster models may be inserted (max. 1500 g) dimensions W
x L x H: 200 mm x 200 mm x
150 mm. Only accessories recommended by CADstar Technology may be used. The scanner is intended to be used in dental laboratories.
Any other use requires the written consent of CADstar Technology. Improper use may endanger persons and damage the scanner. The
product is not an automatically functioning device; it must be loaded manually. Furthermore, any liability and warranty claims are excluded
in the event of non-compliance with the intended use.
1.6 Potential misuse
The information in this documentation describes the properties of the product, but does not guarantee them. No liability and warranty
claims will be accepted for damages caused by:
Improper use of the scanner.
Disregarding the operating instructions.
Unauthorized changes to the scanner.
Improper work on and with the scanner.
Improperly performed repairs.
Unauthorized, improper modifi cation of operating parameters.
Catastrophic events, impacts due to foreign objects and force majeure.
1.7 Intended environment
The limits of use for the scanner are restricted as follows:
To be operated indoors only
Ambient temperature: 5 °C to 35 °C
Relative humidity: up to 31°C 20 % - 80 % at temperatures up to 35°C maximum 20-50% linear decreasing - not suitable for humid
rooms, non-condensing
No fl ickering light (e.g. fl uorescent lighting; moving blinds; shadows from trees in the wind ).
Maximum ambient brightness 500lux. No direct sunlight in the scanning area
Clean and dust free environment
Horizontal installation site on a stable, vibration-absorbing workstation
Maximum installation altitude <3000m above sea level
1.8 Duties of the operator
EN / 4
The swivel arm will move when the scanner is switched on. Do not reach into the danger area as fi nger injuries
may occur.
The swivel arm moves during scanning. Do not reach into the danger area as fi nger injuries may occur.