8
Table 2
MANUFACTURER'S GUIDE AND DECLARATION - ELECTROMAGNETIC EMISSIONS
The scooters are intended for use in the following electromagnetic conditions.
IMMUNITY TESTS
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical Fast Transient/
Burst Immunity
IEC 61000-4-4
Over voltages IEC
61000-4-5
Voltage drop outs, brief
interruptions and variations
of voltage on the power
supply lines. IEC 61000-4-11
Mains frequency
magnetic field
(50/60 Hz) IEC 61000-4-8
Note: UT is the AC mains voltage prior to application of the test level.
The client or user must ensure that it is used in these conditions.
TEST LEVEL IEC 60601
± 6 kV contact
± 8 kV air
±2 kV for power lines
±1 kV differential mode
±2 kV common mode
<5% UT
(>95% UT drop out)
for 0.5 cycles
40% UT
(60% UT drop out)
for 5 cycles
70% UT
(30% UT drop out)
for 25 cycles
<5% UT
(>95% UT drop out) for 5 s.
3 A/m
CONFORMITY LEVEL
± 6 kV contact
± 8 kV air
±2 kV for power lines
±1 kV differential mode
±2 kV common mode
<5% UT
(>95% UT drop out)
for 0.5 cycles
40% UT
(60% UT drop out)
for 5 cycles
70% UT
(30% UT drop out)
for 25 cycles
<5% UT
(>95% UT drop
out) for 5 s.
Magnetic fields at the mains
3 A/m
ELECTROMAGNETIC
ENVIRONMENT – GUIDE
Information for using the
device and charging the
battery. The floor must be
wood, concrete or ceramic.
If the floor is covered
with a synthetic material,
the relative humidity
must be at least 30%.
The quality of the mains
supply must be that of
a typical commercial or
hospital environment.
The quality of the mains
supply must be that of
a typical commercial or
hospital environment.
The quality of the mains
supply must be that of a
typical commercial hospital
when charging the battery.
frequency should have the
typical characteristics of
a commercial or hospital
environment. (Applies both
to using the device and
charging the battery).