Descargar Imprimir esta página

Masimo Rad-G Instrucciones De Uso página 3

Battery replacement kit
Ocultar thumbs Ver también para Rad-G:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 11
Rad-G®
Battery Replacement Kit
The Rad-G Li-Ion Battery Replacement Kit PN 5148 includes the following:
Battery - 27058
CAUTIONS
Only use the battery (PN 27058) included in the kit. Use of other batteries may affect performance.
INSTRUCTIONS
1. Ensure the Rad-G is turned off. Press and hold the power button until the device beeps once and shuts off.
2. Loosen the six (6) screws using a small Philips screwdriver on the back cover (Fig. 1) and remove the cover.
3. Disconnect the battery connector (Fig. 2) by gently tugging on the male six pin connector and remove the battery.
4. Place the new battery in the Rad-G device and connect the male six pin connector into the appropriate female pin connector port.
Note: Route the battery wiring exactly as the original wiring was routed to prevent the wiring from being pinched between the device and
the back cover.
5. Install the back cover. Install and tighten the screws using a small Philips screwdriver.
6. Charge the device and power on to verify battery has been properly installed.
ENVIRONMENTAL
Storage Humidity
Operating Temperature
Storage Temperature
CAUTION: FEDERAL LAW (U.S.A.) RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN.
If you encounter any serious incident with product, please notify the competent authority in your country and the manufacturer.
The following symbols may appear on the product or product labeling:
SYMBOL
Follow instructions for use
Consult instructions for use
Manufacturer
Date of manufacture
YYYY-MM-DD
Use -by
YYYY-MM-DD
Do not re-use/Single patient use only
Non-Sterile
LATEX
Not made with natural rubber latex
Single patient - multiple use
PCX-2108A
02/13
Distributor
Caution: Federal law (USA) restricts this device to
sale by or on the order of a physician
D I R E C T I O N S F O R U S E
10% – 80% non-condensing
0°C – 45°C (32°F – 113°F) at ambient humidity
1 Month -10°C – +55°C (14°F – 131°F)
6 Month -10°C – +45°C (14°F – 113°F)
12 Month -10°C – +25°C (14°F – 77°F)
DEFINITION
SYMBOL
Separate collection for electrical and electronic
equipment (WEEE).
Lot code
Catalogue number (model number)
Masimo reference number
Fragile, handle with care
Greater than
Less than
Caution
Unique device identifier
Medical device
Keep dry
3
DEFINITION
11620B-eIFU-0722
en

Publicidad

loading