Foreword
The non contact Infrared body thermometer operating Instructions intend to provide
the necessary information for proper operation of HTD8813, thermometer model.
Only body mode was reviewed and certi�ied by noti�ied body.
General knowledge of Infrared thermometer and an understanding of the features and
functions of the HTD8813 thermometer model are prerequisites for proper use.
The non contact infrared body thermometer is a medical device, and can be used
repeatedly, which using life is 5 years.
Please read the manual �irst before using it, if not fully understand the usages, please
stop using the thermometer.
Do not operate any of the models HTD8813 thermometer without completely
reading and understanding these instructions.
Notice
Purchase or possession of this device does not carry any express or implied license to
use with replacement parts which would, alone or in combination with this device, fall
within the scope of one of the relating patents.
For further information contact:
Hetaida Technology Co., Ltd.
Room 801, 802, 803, 804,901,
2# Building Scienti�ic Research Center, Songhu
Intelligent Valley, No.6 Minfu Road, Liaobu Town, Dongguan City, Guangdong Province,
P.R.China
Tel:+860769-83326886
Contact: Tom. Chen
E-mail: tomchen@hetaida.com.cn
Catalogue
Safety Information
1- Overview
2- Operation
3-Troubleshooting
4-Replacing the Battery
5-Cleaning, Care and Storage
6-Disposal
7-Warranty
8-EMC Declaration
9-Body temperature
Catálogo
Información de seguridad
1-Descripción general
2-Funcionamiento
3-Solución de problemas
4-Sustitución de la batería
5-Limpieza, cuidado y almacenamiento
6-Eliminación
7-Garantía
8-Declaración FCC (CEM)
9-Temperatura corporal
Share Info Consultant Service LLC Repräsentanzbüro
Heerdter Lohweg 83, 40549 Düsseldorf
Tel: 0049 1767 0057022
Fax:0049 1767 0057022
Email: eu-rep@share-info.com; info@share-info.com
Safety Information
This device may only be used for the purposes described in these instructions. The
manufacturer cannot be held liable for damage caused by incorrect application.
The noncontactinfrared body thermometeris designed to minimize the possibility of
hazards from errors in the software program by following sound and lightengineering
design processes, Risk Analysis and Software Validation.
Warning
Warnings are identi�ied by the WARNING symbol shown above.
The NonContact Infrared Body Thermometer is to be operated by consumers in the
home setting and primary care setting as screening tool. This manual, accessories,
Directions for Use, all precautionary information, and speci�ications should be read
before use.
This product is designed to measure human body temperature on the forehead. Do
not use it for any other purpose.
This product is intended in the home setting and primary care setting as screening
tool.
3
5
10
14
15
16
16
17
18
22
26
28
33
37
38
39
39
40
41
45

Publicidad

loading

Resumen de contenidos para Hetaida HTD8813

  • Página 1 General knowledge of Infrared thermometer and an understanding of the features and Email: eu-rep@share-info.com; info@share-info.com functions of the HTD8813 thermometer model are prerequisites for proper use. Safety Information The non contact infrared body thermometer is a medical device, and can be used repeatedly, which using life is 5 years.
  • Página 2 Intended Use The HeTaiDa Non Contact Infrared Body Thermometers are designed to be used for intermittent measurement and monitoring of human body temperature by consumers in the home setting and primary care setting as screening tool.
  • Página 3 Technical Speci�ications Dimension and Weighting Weight (without batteries) Size L:138mm X W:95mm X H:40mm Compliance Item Compliant with Equipment classi�ication Safety Standards: EN 60601-1: 2006+A1:2013, EN 60601-1-2: 2015 Type of protection Internally powered equipment (on battery power) Degree of protection Non Applied part Front panel and case labeling EN ISO15223-1:2016...
  • Página 4 Green means ready for next measurement. When 37.4℃-38.0℃, it's yellow ,means HTD8813 as below step: slight fever warning. Please pay attention to body temperature. When the body 1)Unit Set-F1 temperature is above 38.1℃, it's red , means high fever. Please take action to cool down or go for a doctor.
  • Página 5 Enquire about the options for environment-friendly and appropriate disposal. Take local regulations into account. 2. Please report to HeTaiDa Technology Co., Ltd. if any unexpected operation or events occur. 3. The thermometer is initially calibrated at the time of manufacture. If this thermometer is used according to the use instructions, periodic re-adjustment is not required.
  • Página 6 Standard (E1965-98) Except of clause 5.2.2. It displays subject’s temperature over a range of 34.0~43.0℃. Full responsibility for the conformance of this product to the standard is assumed by(Hetaida Technology Co., Ltd. Add: Room 801, 802, 803, 804, 901, 2# Building Scienti�ic Research Center, Songhu Intelligent Valley, No.6 Minfu...
  • Página 7 Termómetro de infrarrojos corporal sin contacto favor deje de utilizar el termómetro. Manual de usuario No utilice ninguno de los modelos de termómetro HTD8813 sin haber leído y comprendido completamente estas instrucciones. AVISO La compra o posesión de este dispositivo no conlleva ninguna licencia expresa o implícita para su uso con piezas de repuesto que, solas o en combinación con este...
  • Página 8 12. Indicador de encendido/apagado de la voz 6. Carcasa de plástico 13. Fahrenheit/Celsius Los Termómetros de infrarrojos corporal sin contacto de Hetaida están diseñados 7. Modo super�icial 14. Número de memoria para ser utilizados para la medición intermitente y el control de la temperatura del 15.
  • Página 9 La última temperatura registrada antes de que el termómetro se apague se almacenará en la memoria. Hasta 50 para HTD8813. Para los modelos HTD8813 como se describe en los pasos siguientes: 1) En el estado de arranque o apagado, pulse brevemente el botón MEMO para ver el historial de valores medidos.
  • Página 10 3- Solución de problemas HTD8813 como se describe en los pasos siguientes: 1) Ajuste de la función-F1 En el modo de arranque . Pulse prolongadamente el botón SET para entrar en F1, MENSAJE SITUACIÓN SOLUCIÓN pulse el botón "MODE" o "MEMO "para cambiar las unidades de temperatura Celsius y Fahrenheit, pulse el botón SET para con�irmar los ajustes de las unidades (el valor...
  • Página 11 Prueba de emisión Cumplimiento Emisiones de RF CISPR 11 Grupo 1 2. Informe a Hetaida Technology Co., Ltd. si se produce cualquier operación o evento inesperado. Emisión de RF CISPR 11 Clase B 3. El termómetro se calibra inicialmente en el momento de su fabricación. Si este Emisiones armónicas IEC 61000-3-2...
  • Página 12 34,0~43,0℃. Toda la responsabilidad de la conformidad de LTE Banda 7 este producto con la norma es asumida por (Hetaida Technology Co., Ltd.). 5240 Dirección: Room 801, 802, 803, 804, 901, 2# Building Scienti�ic Research Center, 5100- WLAN 802.11 a/n...