Technical Speci�ications
2- Operation
2.1Battery installation
Caution: The Non Contact Infrared Body Thermometer does not operate with dead
batteries and does not input outer power. Install new batteries.
1)Pull the battery downward, toward the bottom of the Non Contact Infrared Body
Thermometer, and remove the battery access door;
2)Insert two pieces AAA size batteries according to the "+" and "-";
3)Close the battery cover.
2.2How to Operate
Before Applying the Thermometer
Be sure to read and understand all warnings listed of the instructions before use.
The thermometer is aligned with the middle of the forehead to measure body
temperature (between the eyebrows above) and keep the vertical distance, press
the On/measure button, temperature display immediately, see �igure 2.
Figure 2:Measuring position and distance
When the ambient temperature of the thermometer changes too much, such as
moving the Thermometer from one place of lower temperature to another place of
higher temperature, Allow the thermometer to remain in a room for 30 minutes
where the temperature is between15℃ to 40℃.
Dimension and Weighting
Weight (without batteries)
90g
Size
L:138mm X W:95mm X H:40mm
Compliance
Item
Compliant with
Equipment classi�ication
Safety Standards: EN 60601-1: 2006+A1:2013, EN 60601-1-2: 2015
Type of protection
Internally powered equipment (on battery power)
Degree of protection
Non Applied part
Front panel and case labeling
EN ISO15223-1:2016
Temperature
EN ISO80601-2-56:2017
Home healthcare environment
EN 60601-1-11:2015
Calculated values of the indicators according to ISO 80601-2-56
Forehead mode:
Group A1
Bias
0.015
Standard Deviation
0.123
Limits of Agreement
0.245
Clinical Repeatability(for All Group)
Note: the above value is calculated from clinical data of HTD8818A.
Safety classi�ication of ME EQUIPMENT
Protection against electric shock
Applied part
Protection against harmful ingress of water or particulate matter
Mode of operation
Note: Not intended to be sterilized. Not for use in an OXYGEN RICH ENVIRONMENT
The ambient temperature around the test person should be stable, should keep
away from the larger �low fan, air-conditioning vents and so on.
When people moving from one place of lower temperature to another place of
higher temperature, should at least remain in the test environment more than 5
minutes, to be consistent with the ambient temperature after the re-measure-
ment.
Wait at least 1 second for the next measurement. If the continuous measurement
of �ive times, it is recommended to wait at least 30 seconds and then continue
measurement.
You cannot use the thermometer in place where the sun is strong.
If for some reason the low forehead temperature measurement can try to align
behind the ears. See �igure 3.
Figure 3:Align behind the ears to measurement
General Setup and Use
Start measuring
1.Turn on the thermometer by pressing the On/measure button. The thermometer
will perform self-test with all segments displayed for 2 seconds.
2.Align staff forehead to keep the distance, and then press the On/measure button
to start the measurement, read the data.
Note: 1) After full display over, you will hear a rattle or "beep beep beep" three times,
which means that the measurements have been completed, while the target value of
the measured temperature is displayed on the LCD, while backlit display according to
the appropriate setting among the three colors red, green, yellow one of. And the
Group A2
Group B
Group C
0.000
-0.042
-0.040
0.121
0.121
0.137
0.243
0.243
0.274
0.071
Internally powered ME equipment
Non Applied part
IP22
Continuous operation