Maintenance And Cleaning - WIKA DiwiTherm TR75 Manual De Instrucciones

Ocultar thumbs Ver también para DiwiTherm TR75:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6. Commissioning, operation / 7. Maintenance and cleaning
WIKA
Designation
Order no.
per DIN 7603 Form C 14 x 18 x 2 -CuFA
11349981
11349990
per DIN 7603 Form C 18 x 22 x 2 -CuFA
11350008
per DIN 7603 Form C 21 x 26 x 2 -CuFA
11350016
per DIN 7603 Form C 27 x 32 x 2.5 -CuFA
11367416
per DIN 7603 Form C 20 x 24 x 2 -CuFA
1248278
per DIN 7603 D21.2 x D25.9 x 1.5 -Al
3153134
per DIN 7603 Form C D14.2 x D17.9 x 2 -StFA
3361485
per DIN 7603 Form C D33.3 x D38.9 x 2.5 -StFA
11355352
D15 x D21 x 3 -NP-Univ
1605933
D15 x D21 x 1 -NP-Univ
Legend:
CuFA = Copper, max. 45 HB
Al
= Aluminium Al99; F11, 32 to 45 HB
StFA = Soft iron, 80 to 95 HB
6.2 Tapered threads (NPT)
It should be checked whether it may be necessary to seal them additionally with PTFE tape
or hemp. The threads must be lubricated with a suitable lubricant before fitting.
Tighten the threads by hand when carrying out the final assembly on the plant. This will
correspond to the delivery status of the premounted components. The final tightening and
sealing must be made with a spanner (1.5 to 3 rotations).

7. Maintenance and cleaning

7.1 Maintenance
The DiwiTherm
thermometer is maintenance-free.
®
The indicator should be checked once or twice every year.
To do this, the instrument must be disconnected from the process and checked using a
temperature calibrator.
DiwiTherm
model TR75: battery powered (at least 10 years service life)
®
The battery is not replaceable!
Repairs must only be carried out by the manufacturer.
WIKA operating instructions DiwiTherm
a
; filled with asbestos-free sealing material
b
a
; filled with asbestos-free sealing material
, model TR75
®
Suitable for threads
G ¼, M14 x 1.5
M18 x 1.5, G ⅜
G ½, M20 x 1.5
G ¾, M27 x 2
M20 x 1.5
G ½, M20 x 1.5
G ¼, M14 x 1.5
G 1
M24 x 1.5
M24 x 1.5
EN