WIKA SB15 Manual De Instrucciones página 16

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

5. Commissioning, operation
4. Connect the stranded wire according to the pin assignment indicated on the product
label. For this, observe the numbering of connection terminal on the rear of the case.
The screw terminal for the connection of the protective conductor is located at the output
EN
of the measuring line.
5. Re-mount the protection cap and individual components.
5.5 Indicator check
Indicator checks should only be carried out in comparison with a more accurate instrument
or, if possible, with a calibrated instrument. The temperature during the check must remain
constant. Fluctuating temperatures can lead to reading errors caused by the different
response times of the probes.
Before checking the indication, the minimum waiting time of 5 minutes must be observed
(see chapter 5.1 "Mounting the probe").
16
WIKA operating instructions models SB15, SW15

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Sw15

Tabla de contenido