TROUBLESHOOTING
If any abnormality arises during use, please check the following points.
POSSIBLE
SYMPTOMS
CAUSES
Unit does not turn on
Batteries have run
when the START/STOP
down.
button is pushed.
Battery polarities (+/-)
have been positioned
incorrectly.
EE measurement error
The UltraSoft™ upper
symbol shown on
arm cuff has been
display or the blood
positioned on the arm
incorrectly.
pressure value is
displayed excessively
Did you talk or move
low (or high).
during measurement?
Shaking of the arm with
the cuff on.
E1 error symbol shown
Air circuit abnormality.
on Display.
Cuff tube may not be
plugged into monitor
correctly.
E2 error symbol shown
Inflation pressure
on Display.
exceeding 300 mmHg.
E3 error symbol shown
Error determining
on display.
measurement data.
EP error symbol shown
System Error.
on display.
Note: If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair the
unit by yourself. Contact information for HoMedics Consumer Relations
Department can be found on the warranty page.
CORRECTION
Replace them with four new
"AAA" alkaline batteries.
Re-insert the batteries in the
correct positions. See page 11.
Re-wrap the cuff properly so
that it is positioned correctly.
Take new measurement.
Keep arm steady during
measurement. Measure
again. Refer to "Measurement
Procedure" instructions, see
page 17.
Check cuff connection.
See page 15. Take new
measurement.
Switch the unit off and then
turn back on. Take new
measurement.
Re-wrap the cuff properly so that
it is positioned correctly. Take
new measurement.
Remove batteries. After 1
minute, re-insert the batteries.
Take new measurement.
31
SPECIFICATIONS
4 1.5V DC "AAA" alkaline batteries
Power Source:
or 6V DC AC adapter
Measurement Method: Oscillometric
Pressure: 40~280 millimeters mercury (mmHg)
Measurement Range:
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Pump driven
Inflation:
Automatic Pressure release valve
Deflation:
Memory Capacity:
100 memories for each User (200 total)
1 minute after last button operation
Auto-shut-off:
Temperature: 41°F~104°F (5°C ~40°C)
Operation
Environment:
Humidity: 15% ~ 93% RH
Storage/Transportation
Temperature: -13°F~158°F (-25°C~70°C)
Environment:
Humidity: Less Than 93% RH
Weight:
0.58 lbs (265 g) (without batteries)
Arm Circumference:
UltraSoft
4.49"(L) x 5.63"(W) x 1.79"(H)
Dimensions:
114mm(L) x 143mm(W) x 45.5mm(H)
6V DC AC adapter, 1 arm cuff with tube,
Accessories:
intruction manual, 4 "AAA" alkaline batteries
Internally Powered equipment
Type BF Equipment
IP22-Degrees of protection provided by
enclosures
Not suitable for use in presence of flammable
anesthetic mixture with air, oxygen or
nitrous oxide.
Continuous operation with short-time loading.
Follow instructions for use.
Note: These specifications are subject to change without notice.
Upper Arm Cuff size: 9"–17" (23-43 cm)
TM