Descargar Imprimir esta página

Medisana HGC Instrucciones De Manejo página 27

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 55
51237_HGC_WL_NEUSS.qxd:Blutdruckmessgerät
6.1
Troubleshooting
6.2
Cleaning and
maintenance
The battery change symbol
• The batteries are too low or empty. Replace both batteries with new 1.5 V
AAA LRO3 batteries.
Erratic measurements being displayed
• Cuff incorrectly fitted or not tight enough. Fit cuff correctly and tighten.
• You have changed position or the cuff has moved during the measurement.
Keep still and do not move your arm or the cuff.
• You have an irregular heartbeat. Even if your heartbeat is only slightly
irregular, repeat the measurement. The unit is not suitable for people with
an extremely irregular heartbeat.
• You have spoken during the measurement, you are nervous or you are
excited. Do not speak and breathe deeply two or three times in order to
relax.
• You are not in the corrected seated position. Please assume the correct
position when taking the measurement.
• A problem occurred during inflation or during the measurement. Repeat
the measurement.
"EE" appears in the display
• The unit has not been inflated enough. Repeat the measurement
• A problem occurred during inflation or during the measurement. Repeat
the measurement.
No display or display locked
• No batteries have been inserted, batteries wrongly inserted or batteries are
completely empty. Insert batteries, check battery position or replace with
new batteries.
• A malfunction or fault has occurred. Briefly remove the batteries and then
replace them.
Please contact the service centre if the error message appears again
when the measurement is repeated.
• Remove the batteries before cleaning the unit.
• Never use strong detergents or hard brushes.
• Clean the unit using a soft cloth lightly moistened with a mild soapy
solution. Never use abrasive cleaning agents or alcohol.
• Do not allow any water to get into the unit. Do not use the unit again until
it is completely dry.
• Remove the batteries from the unit if you do not wish to use it for an
extended period of time, otherwise there is a risk of battery leakage.
• Do not expose the unit to direct sunlight, and protect it from dust and
moisture.
• Only inflate the cuff when it is in position around the upper arm.
• The unit is best kept in the storage box included in your purchase and should
preferably be stored in a clean dry place.
22.07.2010
4:34 Uhr
6 Miscellaneous
appears in the display
Seite 21
GB
21

Publicidad

loading

Este manual también es adecuado para:

51237