6 - CLEANING, SANITISATION, DISINFECTION
Before and after each use, the device and the accessories must be
cleaned and disinfected properly as described below. If this is not done,
some microorganisms may collect in the device, causing the risk of in-
fection. Do not use alcohol or other solvents to clean the device.
6.1. CLEANING Accessories and Removable Upper Part (2):
PUSH
5
5
4
4
3
3
6.1.1. WASHING: Wash the components 2-2G-
3-4-5 with lukewarm drinking water with dish
washing soap (non-abrasive), then rinse thor-
oughly with a jet of lukewarm drinking water
to remove any detergent residue.
6.1.2. DRYING
6.1.2.a. After washing and rinsing all the com-
ponents, shake them to remove excess water
and reassemble them, as described in the fol-
lowing paragraph 7. REASSEMBLY OF THE
DEVICE. At this point, to eliminate the resid-
ual water that has deposited in the nebuliser
head (2G), press Button 1B (side figure) and
wait for the device stop to nebulizing. Turn
off the unit. This operation is very important,
because it prevents the deposit of limestone
in the micropores of the Nebuliser Head (2G)
which could affect proper nebulisation of the
drug in subsequent therapeutic applications.
6.1.2.b. To complete the drying of all the compo-
2
2
nents, leave them out and in a dry place (e.g.
not in the bathroom) as into the side figure.
2G
2G
WARNING
Do not clean the Mesh (2G) using brushes or
other instruments: this may cause irreparable
damage.