PROBLEM
The device
switches on but
does not nebulise
or nebulises little.
The device
switches off after a
few seconds (blue
light).
The device
switches
off
The START AND
STOP Button (1B)
shows a flash-
ing red light and
the device does
not nebulise or
switches off.
The device does
not switch on.
If after checking the aforementioned conditions, the unit still does not nebulise, we rec-
ommend that you contact the dealer or an authorised Flaem service centre.
TROUBLE-SHOOTING
CAUSE
• Malfunction of the
Mesh (2G)
• The Nebuliser head
(2G) is broken
• No medicine supply
in the nebulising area
• The nebulising speed
depends on the type
of medicinal product
used
• No electrical contact
between the Remov-
a b l e M e d i c a t i o n
chamber (2) and the
Main Unit (1)
• It's been 5 minutes
since device switch-
ing on
• Batteries are used
up or low
• No electric power
supply
SOLUTION
• Soak the Nebulising head
(2G) in a solution of water
(50%) and white wine vinegar
(50%), rinse and fit back on as
described in paragraph 6.3
SANITISING
• Replace it with a new Nebuli-
ser head (2G)
• Verify that the small hole in the
centre of the Medication cup
cover (2D) is not obstructed.
If so clean it with a needle
• The inhaling treatment time
depends on the type of
medicinal product and on the
patient's inhaling ability
• Check that the Removable
Medication chamber (2)
is properly hooked to Main
Unit (1)
• Turn on the unit again
• Replace the batteries if they
are alkaline.
• Recharge the Batteries if they
are rechargeable Ni-MH or
Lithium type.
• Verify that Batteries are insert-
ed with the correct polarity.
• Used up or low Batteries.
Replace or recharge.
16