6th press: high voltage limit – the last 2 digits flash
- use the
To programme the parameters of the next comparator, repeat steps 3 & 4 of the
procedure above.
To exit from the parameter set mode, press the SET key.
5.3

COMPARATOR RESULTS

Interpreting the comparator results :
Resistance
Voltage
Low value
high voltage
limit
High value
5.4

ACTIVATING / DEACTIVATING THE COMPARATOR FUNCTION

1.
press SET to activate the comparator function. COMP is displayed, indicating
that the comparison will be made once the measurement has been carried out.
2.
the comparator used is the last one selected before the instrument was last
switched off. To select another one from among the 99 possible, press
and
.
3.
to activate the warning beep accompanying the WARNING and FAIL ratings of
the comparator, press
deactivate it, press
4.
to deactivate the comparator function, press SET again.
&
keys to set the values of these 2 digits
Low resistance limit
Low value
WARNING LED +
beep
Warning !
PASS LED
Battery OK !
. This same symbol appears on the display unit. To
again.
Battery Tester C.A 6630
High resistance limit
intermediate value
WARNING LED +
beep
Warning !
WARNING LED +
beep
Warning !
33
high value
FAIL LED + beep
Battery Not OK!
FAIL LED + beep
Battery Not OK!