INSTRUCTION MANUAL CODE 80197 REV D Jul/2012
Programs
Step
Sterilizing temperature
Sterilizing pressure
Class
Pre-Heating
Vacuum time
Admission time
Sterilizing time
Drying time
Quick drying time
Total time
Usage (General)
J
.P. SELECTA s.a.
3.3 Bowie Dick test
Use this test to validate the autoclave for wrapped and porous loads sterilization.
3.4 Vacuum test
Use this test to check the vacuum pump and the chamber leakage for a defi ned time.
Vacuum test is recommended to be done everyday before starting the batch steri-
lization.
Ensure that the vacuum gauge in the upper left corner of the screen reaches at least
a value of -0.7 BAR.
B Cycle
18min B cycle
134
134
2.10
2.10
B
B
0-7 min
0-7 min
10 min
10 min
5 min
5 min
4 min
18 min
10 min
10 min
5 min
5 min
36 (31) min
50 (45) min
Class B
Prions
3.5 Guidelines to get test an effi cient sterilization
The material to be sterilized needs to be clean, free from any encrusted
or residual material, wash it in a good detergent and distilled water,
then rinse with distilled water.
Do not overload the trays, left a free space of 1 to 2 cm between goods
to allow a free steam circulation and to ease drying.
3.5.1 Sterilization of instruments in bulk
Place the instruments on a tray with water repellent paper leaving
a space between each instrument. Try to keep similar type material
instruments together on a tray.
3.5.2 Sterilization of instruments in pouches or bags
Place only one instrument per bag.
Place the bags on water repellent paper in a vertical position so that
they do not touch each other or the wall of the autoclave. Better use
a rack.
3.5.3 Tubing
Tubing need to be clean and drained. Place tubing so that the openings
on both ends are open. Make sure that the tubes are not touching the
autoclave wall, blocked or twisted and that nothing can block the inside.
3.5.4 Flasks
Never put hermetically sealed fl asks in the autoclave.
Place the contents open end down to prevent any residual water being
left inside.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(It can be modifi ed without notice)
Unwrapped cycle
Wrapped 121ºC
134
121
2.10
1.10
S
B
0-7 min
0-7 min
4 min
10 min
5 min
5 min
4 min
16 min
10 min
10 min
5 min
5 min
30 (25) min
48 (43) min
Solids loads
Rubber&plastic
Page: 16
BDTest
Vacuum test
134
--
-
< -0.8
B
--
0-7 min
--
10 min
4 min
5 min
--
3.5 min
Keep P. 10min
1 min
--
--
--
19.5 min
15 min
OQ*
OQ*
*OQ: Operation qualifi cation