INSTRUCTION MANUAL CODE 80197 REV D Jul/2012
0.05 bar 83.5 ºC
J
.P. SELECTA s.a.
6.4 System setup
For system setup, time, date, printer, etc. the autoclave must not be running a pro-
gram (Menu state)
•
Press "MENU/C" for 3 seconds to accede SETUP menu.
Use keys «STEP » o «QUICK » to select among the menu options.
Use keys «Menu/C» o «OK» to reject or validate an option.
•
Select «DATE» to enter the date.
•
Select «TIME» to enter the time.
•
Select «PRINTER» to activate the printer report.
•
Select «Time setup» to allow the sterilization time extension from 0 to 3 minutes
added.
•
Select «Temperature setup» to adjust temperature reading. Value from -2 to +3ºC.
Note:
Sterilization time modifi cation must be properly validated for each processed load.
Note:
Do not modify the temperature offset if you do not have a reliable means to measure
the inner chamber temperature.
Any modifi cation of these two parameters can cause an ineffi cient or defective load
sterilization.
6.5 Sterilization end
After sterilization, the autoclave depressurizes the chamber, then you can open the
door.
When the chamber does not have dangerous pressure inside (< 0.05 bar)
shows this situation on the screen and unblocks the electrical lock.
If the printer is attached and ON, printed report is produced.
Use provided holder to remove the hot trays from the chamber.
Caution: when unloading material from the autoclave.
Hazard of burns to be in contact with hot surfaces.
Temperature in this area can exceed 100ºC.
Use the tray holder to load or unload the tray with instruments in it, in order to avoid
burning injuries.