Descargar Imprimir esta página

Vestfrost VLS 026 SDD Instrucciones Para El Uso página 13

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38
Trouble-shooting
Fault
Compressor is not run-
ning, and the ice packs
are not cold
Compressor is running,
and the temperature is
too high
Temperature in VLS
026/056 SDD is too
low
No temperature is
displayed
No light in green diode
during day time
Table 3
Possible cause
Be patient, it is most likely that the
compressor will start within a few
minutes.
The ventilation grill is blocked.
The lid is not closed properly.
The fan in the compressor compart-
ment is blocked or defective.
The temperature in the room in which
the appliance is installed is too high.
To low set point on digital controller
There is not enough light for the solar
sensor.
Switch is turned OFF
Fuse is burned
Diode is defect
Remedy
If this is not the case, check the fol-
lowing:
- Check that power is connected and
that the wire from the solar panel to
the appliance is intact.
- Check the fuse and replace it if
necessary.
- If the above is OK, call technical
supervisor.
Ensure unhindered air circulation.
Ensure that the lid is closed properly.
Check that the fan is running, if not it
should be replaced.
Shield the appliance against direct
sun light and ensure more ventilation
to the room.
Contact technical supervisor for
ajustment.
Turn on the light.
Turn on the switch
Replace the fuse.
Replace diode.
13
GB

Publicidad

loading

Este manual también es adecuado para:

Vls 056 sdd