Preparation
• Directly following each treatment, all parts of the
atomiser [9] and any accessories used must be
cleaned of medicine residues and contamination.
• To do this, dismantle the atomiser [9] into its com-
ponent parts.
• Remove the mouthpiece [12] from the atomiser.
• Dismantle the atomiser by twisting the upper section
anti-clockwise against the medicine container [11].
• Remove the nozzle attachment [10]
from the medicine container [11].
• The nebuliser is later reassembled
in the reverse order.
Cleaning
Important
Before you clean the unit, always switch it off, unplug
it and let it cool off.
The atomiser and any accessories used, such as the
mouthpiece, mask, nozzle attachment, etc., must be
cleaned with hot water.
Carefully dry all components using a soft cloth. Once
all components are fully dry, reassemble the atomiser
and place the components in a dry, sealed container
or carry out disinfection.
When cleaning, ensure that all residues are removed.
Never use any substances that may be poisonous
when in contact with the skin or mucous membranes,
or when swallowed or inhaled.
Clean the device using a damp cloth, to which you can
apply a little detergent if necessary.
Do not use abrasive detergents and never immerse
the unit in water.
Important
• Ensure that no water penetrates inside the unit!
• Do not put the device or accessories in a dishwasher.
• Do not touch the unit with wet hands while it is
plugged in; do not allow any water to be sprayed
onto the unit. The unit must be operated only when
it is completely dry.
• Do not spray any liquids into the ventilation slots. Any
liquid that penetrates into the unit can damage the
electrical parts or other components of the nebuliser
and impair the function of the device.
Condensation, care of the hose
Depending on the environmental conditions, con-
densation may collect in the hose. To prevent against
bacterial growth and to ensure effective therapeutic
use, this moisture must always be removed. To remove
moisture from the hose, proceed as follows:
• Remove the hose [8] from the atomiser [9].
• Leave the hose plugged into the nebuliser [6].
• Run the nebuliser until all the moisture has been dis-
placed by the flow of air.
• In the event of significant contamination, replace the
hose.
Disinfection
Follow the steps listed below carefully in order to disin-
fect your atomiser and the accessories. It is advisable
to disinfect the individual components after the last
time they are used each day at the latest.
10
• First clean the atomiser and the accessories as de-
scribed under "Cleaning".
11
• The atomiser should be taken apart and placed to-
gether with the mouthpiece and nosepiece in boiling
hot water for at least 10 minutes. Avoid the parts
coming into contact with the hot base of the pot.
• You can also disinfect the atomiser, the mouthpiece
and the nosepiece with a standard vaporizer. Ob-
serve the manufacturer's instructions for use for the
vaporizer when doing this.
Note
The face masks and air hose must not be placed in
hot water!
• Carefully dry all components using a soft cloth.
• We recommend disinfecting the masks using a retail
disinfectant.
• Once all components are fully dry, reassemble the
atomiser and place the components in a dry, sealed
container.
Note
Ensure that all parts are thoroughly dried during, and
in particular after cleaning, otherwise there is an in-
creased risk of bacterial growth.
Drying
• Place the individual components on a dry, clean and
absorbent surface and allow to dry completely (for
at least 4 hours).
Material resistance
• As with all plastic components, frequent use and
cleaning of the atomiser and the accessories can
lead to a certain amount of wear. Over time, this can
alter the aerosol properties and may eventually affect
the efficiency of therapy. We therefore recommend
that you replace the atomiser once a year.