6. INTRODUCTION TO PANEL
SETTING ENTRY:
Press the
button on the operating panel in the state of normal weighing.
If F1.14 = 0, you can set all the parameters within F1~F5.
If F1.14 = 1, you can only set all parameters within F2~F5.
If F1.14 = 1 and you need to set the parameters within F1 menu, you can press the calibration switch button until the F1 menu is
entered.
F1 PARAMETER SETTING OF SCALE
F1.1 Measuring Range
Selectable parameters: 3~200000 (default value: 6)
F1.2 Decimal Places
Selectable parameters: 0 ---- no decimal point
0.0 ---- 1 decimal place
0.00 ---- 2 decimal places
0.000 ---- 3 decimal places (default value)
0.0000 ---- 4 decimal places
F1.3 Number of Divisions
Selectable parameters: 1 (default value), 2, 5, 10, 20, 50
F1.4 Calibration Unit
Selectable parameters: 0 ---- kg (default value)
1 ---- lb
F1.5 Gravitational Acceleration
Selectable parameters: 9.70000~9.99999. Default value = 9.79455.
F1.6 Null-point Calibration
[E-SCL] Keeping empty the scale
Remove the weights on the weighing platform to guarantee the scale is in the empty state. Press the
play [0 cal]. The displayed digits will reduce slowly until the meter displays [00 cal ]. In the end it will display [End] for one second,
which indicates the end of null-point calibration.
clearing key
Gross weight state resets weight. When the scale is in
net weight, dynamic state, saving state and out of reset-
ting range, clearing operation is invalid
ON/OFF key
Short key → start up or print. (only for printer models)
Printing format refers to Appendix1.
Long key → shut down. Turn on/off
USER'S MANUAL ARX
In setting, digit of flicker posi-
tion increases. In adjustment
of display, accumulation is
cleared.
Confirm operation, to save
setting data.
key and the meter will dis-