The table below details the five function options available to you in Setup
mode.
Function
Option No.
(Left Digit) (Right Digit)
Default
0
Settings
Source
Beeper
1
Auto power
2
off
Backlight
3
Calibration
4
Check
The factory default setting for all five functions—including the source of the
default settings—is 0.
To change any or all settings to your choice(s), you must first change the
setting of Option 0 from 0 to 1. Opportunities to change the settings of the other
four functions from their factory defaults are then made available in order, from
Option 0 to Option 4. As each Option is presented, you can change its setting
or leave the existing setting unchanged. In either case, you must confirm the
setting to allow the sequence to continue.
It's important to understand that in Setup mode you have 30 seconds to
confirm the setting of each Option. If you take longer than 30 seconds to
confirm any setting, the MMD950 will automatically exit Setup mode and enter
Measurement mode.
8
Setting
Action
0
Loads factory default settings
1
Loads user's default settings
Backlight comes on for 30 seconds
0
Programs beeper to beep faster the higher the
moisture level above 17%WME (17 in REL mode)
1
Programs beeper to beep faster the higher the
moisture level
2
Programs beeper to sound when switching
between measurement modes
3
Disables beeper. Removes
0
Disables auto power off function
1
Sets auto power off interval to 3 minutes
2
Sets auto power off interval to 5 minutes
3
Sets auto power off interval to 10 minutes
0
Disables display backlight
1
Enables display backlight
0
Enables manual calibration checking by user
1
Programs meter to automatically check its
calibration each time it is powered on
icon from display