9. Specifications
Operating conditions
Operating conditions
Vibration resistance
Shock resistance
Single shock loads
Continuous shock
Permissible temperature ranges
Ambient temperature
Storage temperature
Permissible air humidity
Ingress protection
1)
M12 x 1 plastic
M12 x 1 metal
Cable outlet plastic
Cable outlet metal
Field case
Electrical safety
Model GD-20-D
Model GD-20-A
1) The respective IP protection rating applies to electrical ouputs with plug-type design when plugged in or with cable/bling plug, respectively. The mating connector
must be suitable for the required ingress protection.
WIKA operating instructions gas density sensor, model GD-20
All manuals and user guides at all-guides.com
5 g, 15 ... 2,000 Hz, for versions with field case
■
20 g, 30 ... 2,000 Hz, for versions without field case
■
500 g (1.4 ms, 1 shock, 3 axes)
100 g (4 ms, 10,000 shocks, 3 axes)
-40 ... +80 °C [-40 ... +176 °F]
-40 ... +70 °C [-40 ... +158 °F]
≤ 95 % r. h. (non-condensing)
IP67, with mating connector
IP67, with mating connector
IP67, with cable
IP67, with cable
IP6k9k, with cable/blind plug
Reverse polarity voltage U
Reverse polarity voltage U
vs. U
DC 30 V
+
-
vs. U
DC 40 V
+
-
EN