Err 7
Err 8
Err 9
Err 10
Err 11
Err 12
Err 13
Err 14
Err 15
Err 16
Err 17
Err 18
Err 19
Err P
Err l
Err--ol--
Err--lo--
Err
FA , L
H
FA , L L
FA , L
Err Ba Lo
Lo bA
8
Percentage Error
Calibration weight error
Unstable Reading
Wireless communiation failure
Communication protocol error
Accumulation error
Lack of unit weight
Lack of sample
Gravity error
Paper error
Tare out of range
Pre-tare error
Initialize zero error
Printer error
Approval setting error
Over range
Underload
Calibration Error
Battery low
Please check input data, must be > 0.5d.
Check the test weights for calibration or linear calibration
Check any air variation, vibration, RF noise and touching some whe-
re. Check the load cell and connecters.
Check wireless settings or change Com settings from the wireless.
Check communication settings.
Max accumulation times 99 / 999 / weight 999999.
Check unit weight entry data, must be >0.5d.
Check counting samples entry, must be >20d.
Check the gravity settings. Gravity range must be 0.9xx~1.0xx
Check the printer paper.
Minus weight or overload. Remove the load and restart scale again.
Check the pre-tare value.
Calibration the scale.
Check the printer settings or connections
Check the PCB jumper settings. Must be connect
jumper pin to K1 (BW series)
Remove the load. Re calibrate
Minus weight, check the platform and restart or calibrate.
Check the test weights & Re calibrate
Re charge battery, check the voltages.
USER'S MANUAL BS-TRI