Descargar Imprimir esta página

ResMed AirSense11 Guia Del Usuario página 15

Publicidad

Idiomas disponibles

Idiomas disponibles

To set up the device:
1. Place the device on a stable level surface.
2. Connect the power cord into the power inlet at the rear of the device. Connect one end of the power
cord into the AC adaptor and the other end into the power outlet. Ensure the device is set up and
connected to power to enable settings to be applied wirelessly to the device if required.
3. Connect the air tubing firmly to the outlet connector at the rear of the device.
4. Open the humidifier tub and fill it with water.
Note: The humidifier tub must be removed from the device before adding water.
If using the HumidAir 11 Standard water tub, use distilled water only
If using the HumidAir 11 Cleanable water tub, use drinking quality water (potable).
Fill the water tub up to the maximum water level mark. The humidifier tub has a maximum capacity of
380 mL.
5. Close the humidifier tub and insert it into the side of the device.
6. Connect the free end of the air tubing firmly onto the assembled mask.
See the mask user guide for detailed information.
Recommended masks for use with this device are listed on ResMed.com.
Notes:
Do not insert any USB cable into the AirSense 11 device or attempt to plug the AC adaptor into a USB
device. This may cause damage to the AirSense 11 device or USB device.
The electrical connector end of the heated air tubing is only compatible with the air outlet at the device
end and should not be fitted to the mask.
Do not use electrically conductive or anti-static air tubing.
6

Publicidad

loading