ELECTROMAGNETIC COMPATIBILITY
GUIDANCE AND MANUFACTURE'S DECLARATION-ELECTROMAGNETIC IMMUNITY
FOR EQUIPMENT AND SYSTEMS THAT ARE NOT LIFE-SUPPORTING
Guidance and manufacture's declaration-electromagnetic immunity
The Pulse Oximeter is tended for use in the electromagnetic environment specified below. The user of
the Pulse Oximeter should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Radiated RF
3V/m
ICE 61000-4-3
80MHz to 2.5GHz
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength
in the location in which The 11-50DP Pulse Oximeter is used exceeds the applicable RF compliance
level above, the 11-50DP Pulse Oximeter should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the
11-50DP Pulse Oximeter.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3V/m.
26 • ENGLISH
Compliance
Electromagnetic environment-guidance
level
Portable and mobile RF communication
equipment should be used no closer to any part
of the 11-50DP Pulse Oximeter, including cables,
than the recommended separation distance
calculated from the equation applicable to the
3V/m
frequency of the transmitter.
recommended separation distance
Where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a should be less than the compliance level in
each frequency range. Interference
may occur in the vicinity of equipment
marked with the following symbol:
x
80MHz to 800MHz
800MHz to 2.5GHz