Compliance With International Standards; Electromagnetic Compatibility - Chauvin Arnoux MA 100 Manual De Instrucciones

Sensores flexibles de corriente ac
Ocultar thumbs Ver también para MA 100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Housing:
Overall dimensions: 120 (pin output) or 103 (BNC output) x 64 x 28mm
Output by two male pins, dia. 4mm, 19mm centres, or by a coaxial cable 40 cm
long terminated by an insulated female BNC connector.
Mass of the instrument: approximately 200g.
Index of protection:
Flammability rating:
Flexible sensor:
Housing:
The flexible coil is resistant to oils and aliphatic hydrocarbons.

3.9. COMPLIANCE wITH INTERNATIONAL STANDARDS

Electrical safety as per IEC 61010‑2‑032 (Ed. 2 of 2003) for type B sensors.
Double insulation
Measurement category: III
Rated voltage: 600V
(1): or 300V in category IV.
(2): or 1000V in category III.

3.10. ELECTROMAGNETIC COMPATIBILITy

The instrument satisfies the EMC and LVD directives required for the CE marking
and product standard IEC 61326‑1 (Ed. 97) + A1 (Ed. 98).
Immunity in industrial environment
Emissions in industrial environment.
IP 50 as per IEC 60529 (Ed. 92)
IK 04 as per IEC 50102 (Ed. 95)
V0 (as per UL 94)
V2 (as per UL 94)
Housing
(1)
Double insulation
Measurement category: IV
Rated voltage: 600V
9
Sensor
(2)

Publicidad

Tabla de contenido
loading

Tabla de contenido