Hygiene and maintenance
WARNING!
Respect all directives, standards and national legislation
governing cleaning, disinfection and sterilisation.
DANGER!
Always wear gloves to avoid the risk of lesions and
infections when checking, cleaning, repairing or extracting
teeth.
CAUTION!
Never immerse the handpiece in disinfectant solutions
or ultrasonic baths! Only use chlorine-free disinfectant
products that have been certified by officially recognised
institutes.
-External disinfection
Clean and disinfect the handpiece after every treatment
to eliminate residues of bodily fluids (e.g. blood, saliva
etc.) and prevent them entering the internal parts of the
handpiece.
1) Clean off all dirt and impurities from the handpiece. Do
not use metal brushes.
2) Clean the handpiece with disinfectant wipes.
3) Blow the handpiece dry with a jet of compressed air.
- Cleaning the spray nozzles
Use the cleaning wire provided to gently remove dirt and
deposits from the spray nozzles. Clean and disinfect the
cleaning wire.