Troubleshooting
Issue
No stimulation
from pads
No stimulation
from electrode
pads. Screen
displays
Stimulation
feels weak
Device beeps
Nothing displayed
on screen
Gel does not
adhere to the skin
Stimulation is felt
more strongly
through one
electrode pad
Buttons are
not responsive
14
Suspected cause
Low battery or
connection error.
Connection error.
The electrode
pads may have
come loose from
the skin. Or the
connecting cable
may have detached
or be damaged.
Poor connection or
worn gel pads
Low battery.
The beep sound
indicates that
battery is running
out
Exhausted battery
Gel pads have
become overused
Poor positioning of
electrode pads
Lock function may
be active
Suggested action
Fully charge Ellune's battery.
Check that the connection cable
is properly connected to the
electrode pads and to your Ellune
device, and that the cable is not
damaged.
Firmly press the electrode pads
against your skin. Check that
the connection cable is properly
connected to the pads and to your
Ellune device. Replace cable if
there are signs of damage.
Firmly press the electrode pads
against your skin. If this does not
improve stimulation, apply brand
new gel pads to the
electrode pads.
Fully charge Ellune's battery
Fully charge Ellune's battery
Replace the gel pads
Ensure the electrode pads are
positioned on either side of your
pain and at the same height on
your body
Press the lock button once to
reactivate the other buttons