600 V
* When input is opened, fluctuating up to 30 digits; measuring range ≥ 1 mV
7.3 Resistance range
No-load voltage approx. 0.3 V, max. test current 0.2 mA
Measuring range
400 Ω
4,000 kΩ
40,00 kΩ
400,0 kΩ
4,000 MΩ
40,00 MΩ
in end area of display only, + 6 digits
*1
in end area of display only, + 3 digits
*2
maximum start-up time until display appears, 20 sec.
*3
7.4 Continuity test
No-load voltage approx. 3 V, max. test current 0.1 mA
The built-in buzzer sounds when resistance is less than 50 Ω - 300 Ω.
7.5 DC current ranges
Measurement accuracy at a temperature of 23 °C ± 3 °C
Measuring range
40.0 A
40.0 - 200.0 A
200.0 - 300.0 A
7.6 AC current ranges
Measurement accuracy at a temperature of 23 °C ± 3 °C. The measuring value
is gained and indicated as effective value (True RMS, AC coupling). Its calibra-
tion is adapted for a sinusoidal curve form. With deviations from this form, the
display value becomes less accurate. The additional error for the following crest
factors is as follows:
crest factor of 1.4 to 2.0: additional error ± 1.0%
crest factor of 2.0 to 2.5: additional error ± 2.5%
crest factor of 2.5 to 3.0: additional error ± 4.0%
Measuring range
0 - 4.0 A
4.00 - 40.00 A
40.0 - 200.0 A
200.0 - 300.0 A
*
measuring range ≥ 0,1 A
05/ 2015
± (1,5 % of reading + 5 digits)
1 V
in frequency range 40 Hz - 500 Hz
Resolution
Measurement accuracy
100 mΩ
± (1,2 % of reading + 6 digits)
1 Ω
± (0,9 % of reading + 3 digit)
10 Ω
± (0,9 % of reading + 3 digit)
100 Ω
± (1,2 % of reading + 3 digit)
1 kΩ
± (1,2 % of reading + 3 digit)*
10 kΩ
± (2,5 % of reading + 5 digit)
Resolution
Measurement accuracy
10 mA
± (1.5 % of reading + 2 digit)
100 mA
± (1.5 % of reading + 2 digit)
100 mA
± (2.0 % of reading + 2 digit)
Resolution
Measurement accuracy
± (1.0 % of reading + 5 digit)*
for frequencies 50 - 60 Hz
10 mA
± (2.0 % of reading + 7 digit)*
for frequencies 40 - 1 kHz
± (1.5 % of reading + 3 digit)*
for frequencies 50 Hz - 60 Hz
10 mA
± (2.5 % of reading + 5 digit)*
for frequencies 40 Hz - 1 kHz
± (1.5 % of reading + 3 digit)
for frequencies 50 Hz - 60 Hz
100 mA
± (2.5 % of reading + 5 digit)
for frequencies 40 Hz - 1 kHz
± (3.0 % of reading + 3 digit)
for frequencies 50 Hz - 60 Hz
100 mA
± (5.0 % of reading + 5 digit)
for frequencies 40 Hz - 1 kHz
BENNING CM 2
600 V
600 V DC voltage
Overload protection
600 V
*1
600 V DC voltage
600 V
*2
600 V DC voltage
600 V
*2
600 V DC voltage
600 V
*2
600 V DC voltage
600 V
2
600 V DC voltage
600 V
*1 *3
600 V DC voltage
Overload protection
400 A
400 A
400 A
Overload protection
400 A
400 A
400 A
400 A
eff
eff
eff
eff
eff
eff
eff