Task selection menu:
6
MENU
1
OK
2
(Table 2 & Table 3)
If you attach a Universal Measurement Module (UMM) to
the communications port (Figure A1 - Item 9), the task
selection menu shows the applicable IDOS options.
Make the necessary selections from each area ( and ).
One task is permitted in each area.
Note: Use the Dual Function area () to do two operations
at the same time. If the area  selection is not necessary,
set this area to off (). This saves the battery power.
Table 2: Menu Options - Task Selections (Dual
Function, Area )
Options
Description
(If applicable)
Input measurement tasks:
mA
Measure ±55 mA
V
Measure ±30 V
mV
Measure ±120 mV
Hz
Measure the frequency (Units: Table 4)
RTD
Measure RTD temperature
Measure RTD resistance or Ω
TC
Measure thermocouple temperature OR mV
Only when an IDOS UMM is attached. An IDOS
measurement task.
Output tasks:
mA
Supply 0 to 24 mA
V
Supply 0 to 12 V
mV
Supply 0 to 120 mV
Hz
Supply an output frequency (Units: Table 4)
RTD
Simulate RTD temperature
Simulate RTD resistance or Ω
TC
Simulate thermocouple temperature OR mV
Table 3: Menu Options - Task Selections (Area )
Options
Description
(If applicable)
White button = A Dual Function is set.
Black button = Dual Function, area  is set to off.
Input measurement tasks:
mA
Measure ±55 mA
V
Measure ±30 V
mA/24V Measure ±55 mA (24V loop power is on)
A switch test.
Only when an IDOS UMM is attached. An IDOS
measurement task.
Copyright 2005 Baker Hughes Company.
4 | DPI880 Instruction Manual–English
Display:
mA output
RTD
V
COM
Hz
0.000
1
mA +
24mA
TC
3W
30V
MAX
2
-mA (24Vo)+
Sk1
Sk2
mA
Vin
24Vo
30V MAX
Sk1 = Edit
Sk2 = Settings
1.7 Set Up the Settings
When the task is set up (Table 2 & Table 3), use the
Settings menu to adjust the input and/or output operation.
Display: Task
mA + IDOS
mA -
4W
0.000
24mA
0.0
10bar g
Sk1
Sk2
Sk1 = Start/Stop
Sk2 = Settings
Menu:
Settings
... Units
(Table 4)
Menu:
Settings
Sk2
(Table 5)
If there is additional data for a menu option, select Settings
() to see the values that are set up. If necessary, adjust
the values.
Table 4: Menu Options - Settings (Input)
Options
Description
(If applicable)
Pressure Units (UPM only). If you select an
... Units
IDOS task (Table 2 & Table 3). Select one of the
fixed units of measurement (psi, mbar ...).
Temperature Units (RTD or TC only). To select
the temperature units (°C or °F).
Frequency Units (Hz only). To select one of
these units:
Hz: Range < 1000 Hz
kHz: Range 0 to 50 kHz
counts/minute (cpm)
counts/hour (cph)
(TC only). Change the measurement operation:
...
Temperature to mV OR
mV to Temperature
(TC only). To select the type of cold junction (CJ)
compensation.
Automatic: The instrument monitors the CJ
temperature and applies the necessary CJ
CJ ...
compensation.
Manual: Measure the CJ temperature and set
the applicable value. The instrument uses this
value to apply the necessary CJ compensation.
Settings selection
1
(If applicable)
mA

IDOS
Display:
4A
5A
... units
bar
MENU
OK
Pa
(Table 4)
4B
5B
MENU
OK
[]
2
3
MENU
OK
6A
7A
MENU
OK
[]