4 Installation
Inspection
In case of damage from transport is noticed, contact your local distributor.
DO NOT USE the device!
Installation
The EN-Tree P must be installed in such a way that its use is not limited by other objects. Keep at least
2.50 m free around the unit.
The EN-Tree P must be attached to the wall with suitable fittings. The EN-Tree P must only be
mounted on a brick or concrete wall. Position the plugs firmly in the centre of a brick in case of a brick
wall. The fixation must be able to resist a horizontal pulling force of 950 N.
In case of installation of 2 EN-Tree P pulley, we advise you to place the centre of the EN-Tree P at
2.5 m from both sides of the corner (figure 1).
Figure 1
Connection of an adapter other than the prescribed charger can adversely affect the safety of the user
and the functioning of the unit, and is therefore not permitted.
The device is not intended for use in so-called "wet rooms" (hydrotherapy rooms).
Page 5 of 16